High Quality Bone Densitometer - AMAIN Automatic Hematology Analyzer AMHA3100 – Amain Detail:
Product Description
AMAIN Automatic Hematology Analyzer AMHA3100 Clinical Chemistry Analyzer With Touch Screen

Image Gallery


Specification
MAIN TECHNICAL SPECIFICATIONS
| Test parameters | WBC 3-part differential counting, 23 parameters (including WBC, RBC, PLT color histogram) |
| Measurement principle | Counting by electrical impedance method, colorimetric method to measure HGB |
| Quality control method | L-J, automatic drawing and printing of quality control charts |
| Sample volume | Trace detection, 10μL peripheral blood or anticoagulant blood, pre-dilution mode 20μL |
| Precision | WBC (white blood cell) CV≤4.0%, RBC (red blood cell) CV≤2.0%, HGB (hemoglobin) CV≤2.0%, PLT (platelet) CV≤8.0%, MCV (average red blood cell volume) CV≤3.0% |
| Accuracy | The allowable relative deviation range: WBC≤±15%, RBC≤±6.0%, HGB≤±6.0%, PLT≤±20.0%, HCT (hematocrit)≤±9.0% |
| Blank count | WBC≤0.5×109/L, RBC≤0.05×1012/L, HGB≤2.0g/L, PLT≤10.0×109/L |
| Carry-over | WBC≤3.5%, RBC≤2.0%, HGB≤2.0%, PLT≤5.0% |
| Linearity deviation | WBC≤±5%, RBC≤±5%, HGB≤±3%, PLT≤±10% |
| Related coefficient | WBC≥0.990, RBC≥0.990, HGB≥0.990, PLT≥0.990 |
| Display | Color LCD touch screen |
| Detection channel | Dual channel |
| Test speed | 35 (or 60) specimens/hour, continuous working for 24 hours |
| Data storage | It can automatically store more than 30,000 groups of complete results (each result has three histograms) |
| Interface | RS232 interface, VGA interface |
| Power supply | 100V-240V; 50/60Hz |
Product Application
INTRODUCTION
The Automatic Hematology Analyzer is an in vitro diagnostic device that used for quantitative analysis of blood cells, and it can realize three classifications of the count results of white blood cells. This analyzer is a clinical inspection instrument for screening. When making a clinical judgment based on the analysis results, the doctor should take the clinical examination results or other test results into consideration. This analyzer is suitable for the detection of white blood cells, red blood cells, platelets, hemoglobin and other parameters and counting of white blood cells into three classifications.

Product detail pictures:






Related Product Guide:
We thinks what clients think, the urgency of urgency to act from the interests of a client position of theory, allowing for greater quality, lessen processing costs, price ranges are much more reasonable, won the new and outdated shoppers the support and affirmation for High Quality Bone Densitometer - AMAIN Automatic Hematology Analyzer AMHA3100 – Amain, The product will supply to all over the world, such as: Belgium , Belgium , Finland , We have more than 200 staff including experienced managers, creative designers, sophisticated engineers and skilled workers. Through hard work of all employees for the past 20 years own company grew stronger and stronger. We always apply the "client first" principle. We also always fulfill all contracts to the point and therefore enjoy excellent reputation and trust among our customers. You are very welcome to personally visit our company.We hope to start a business partnership on the basis of mutual benefit and successful development . For more information please do no hesitate to contact us..